Build Your Online Product Catalogs?
Product Name: |
digital armtype blood pressure monitor
|
Supply Ability: |
|
Related proudcts |
blood pressure monit, sphygmomanometer, ce, |
Specifications |
ARM TYPE X5 |
Price Term: |
|
Port of loading: |
|
Minimum Order |
|
Unit Price: |
|
|
Pneumatic System£º 1.Pressure time: <60s 2.Closed gas path: air leakage should not be greater than 6 mmHg/min.£¨or 0.8 kPa/min£©£¨500 cc TANK£© 3. Air leakage rate: the pressure drop rate of the armband pressure from 250 mmHg to 50 mmHg is (2~6 mmHg/s) 4. Automatic zero reset: Automatic zero reset after each boot. 5. Inflation rate: The pressure in the 200©P container is charged from 0mmHg to 200mmHg (26.7kPa) within 10 seconds. 6. Air tightness of air pump: When the volume of the container is <=100CC, the pressure in the pipeline is 250 mmHg, 150 mmHg, 50 mmHg, and the air leakage is less than 1 mmHg/30s 7. Airtightness of the pipeline system: the leak rate of the entire pipeline system of the sphygmomanometer causes the pressure drop rate to be less than or equal to 6 mmHg/min 8. The air tightness of the quick vent valve: when the volume of the container is <=80CC, the pressure in the pipeline is 250 mmHg, 150 mmHg, 50 mmHg, and the leakage is less than 1 mmHg/10s 9. Bleeding method: slow bleed valve + fast bleed valve |
Company: |
JIANGXI AICARE MEDICAL TECHNOLOGY CO., LTD
|
Contact: |
Ms. Alice Song |
Address: |
South Side of South Ring Road, Le'an County, Fuzhou City, 344300 Jiangxi Province, China |
Postcode: |
344300 |
Tel: |
+8615931189931 |
Fax: |
|
E-mail: |
|
|
|
|